|
21 CFR 900.12(e)(13) Facilities shall
establish and comply with a system specifying procedures to be followed
by the facility for cleaning and disinfecting mammography equipment
after contact with blood or other potentially infectious materials.
This system shall specify the methods for documenting facility
compliance with the infection control procedures established and shall:
§
Comply with all applicable Federal, State, and local regulations
pertaining to infection control; and
§
Comply with the manufacturer's
recommended procedures for the cleaning and disinfection of the
mammography equipment used in the facility; or
§
If adequate manufacturer's
recommendations are not available, comply with generally accepted
guidance on infection control, until such recommendations become
available.
Comments:
The Infection control policy should
include the following information:
§
Name of the disinfectant used
§
Manufacturer recommendations for
disinfectant
§
Copy of disinfectant reference
material, for example: instructions for use
§
Step-by-step instructions on how
to disinfect the mammography unit when contaminated with blood or other
potentially infectious materials, for example: where to spray the
disinfectant (on the cloth), and how long to leave the disinfectant on
the unit, etc.
§
How to disinfect gross
contamination (usually indicated on the disinfectant)
§
How disinfection will be
documented and where (verification of disinfection is an MQSA
requirement)
|