Mobile Units
Equipment Quality Control
Referenced from the Policy Guidance Help System:
Citation:
900.12(e)(7): Mobile Units. The facility shall
verify that mammography units used to produce mammograms at more
than one location meet the requirements in paragraphs (e)(1) through
(e)(6) of this section. In addition, at each examination location,
before any examinations are conducted, the facility shall verify
satisfactory performance of such units using a test method that
establishes the adequacy of the image quality produced by the unit.
Discussion:
Question 1: Does a mobile unit need to include a
film processor?
No. A mobile unit does not need to include a film processor.
However, any processor used for films taken by the unit must be
identified and is subject to MQSA inspections for verification that
the equipment meets quality standards.
Question 2: What would represent an appropriate
method for repeat analysis for my mobile service that consists of
several individually certified facilities? Should there be an
independent analysis for each facility or a combined analysis for
all of my facilities?
The repeat analysis must be facility based and not combined with
your other facilities. The analysis must be done at least quarterly,
and, if possible include at least 250 examinations.
Question 3: How does a facility demonstrate
satisfactory performance for mobile units after they are moved to a
new location?
For those facilities with mobile units, each mammography unit
must be tested after moving to a new examination location and before
examining any patients to verify the adequacy of the image quality
produced by each unit.
As an example of an acceptable test, a phantom image can be taken
in the AEC mode (or the mode used clinically) after the move but
prior to patient examination. This image is then either processed
and evaluated at the mobile unit site (if possible), or processed
off-site and evaluated to verify performance prior to examining
patients. A passing object score for this phantom image will
be accepted as evidence that the unit is performing adequately after
the move and before patient examination.
Another example, for use when processing is not immediately
available, is to (1) for a given kVp, record the mAs resulting from
a phantom exposure (in the AEC mode under typical clinical
conditions or the mode used clinically); (2) compare that mAs to a
standard mAs value previously established as ensuring output
consistency; and (3) if the two readings are within +/- 10%, proceed
with clinical examinations; otherwise take corrective actions to
bring the two values within this limit before proceeding with
clinical examinations. A crucial follow-up to this test by the
facility is to process (using a processor in control) and score the
objects in the phantom image taken in step (1) at the earliest time
available and before batch processing any of the clinical images. If
this phantom image score fails because of any processing problems,
the problems should be corrected prior to processing any of the
clinical images. If the phantom image score fails due to a
non-processor problem, the mobile facility should still process all
the films. Each clinical exam should be evaluated individually to
determine whether any of the patients have to be recalled to have
their images repeated. The entire imaging chain must be checked and
adjusted or repaired prior to further clinical use.
If the facility takes a phantom image as part of its
post-move/pre-examination testing, it needs to document the object
score of the phantom image. The facility needs to keep the written
records of post-move/pre-examination tests for the last 12 months or
since the last inspection, whichever is longer, and the phantom
images for the last 30 days.
Other tests designed by qualified personnel (the medical
physicist should be consulted) could be acceptable but may have to
be evaluated by the inspector on a case-by-case basis.
Question 4: We are a mobile facility with a van
that does not have on-board processing. We have a film-changing room
on the van where we load and unload cassettes with film during the
day. When the van returns to our main office, we batch process the
films. Do we have to perform a semiannual test for darkroom fog in
this film-changing room?
Yes. Since you
use a room to load and unload cassettes with film, you must test
this room for darkroom fog as well as the darkroom for your
processor. Both rooms have the potential to fog films and degrade
image quality.
Question 5: We use FDA’s guidance for mobile
facilities where we produce a phantom image after a move of the
mobile unit and we monitor the mAs. We then process the phantom
image later, prior to processing the mammograms. If we move the
mobile unit more than once per week, do we also have to produce a
weekly phantom image in addition to the phantom produced after each
move?
If you use the mode of operation and/or technique factors used
clinically for a standard breast for the phantom images that you
produce after each move, you do not have to perform an additional
weekly phantom image. However, the image used for the weekly phantom
test must be evaluated for all required density measurements in
addition to the object scores. 21 C.F.R. 900.12(e)(2).
|