| Facility Scorecard Article Posted to the FDA
Mammography Website 9/27/06
Establishing and/or Changing the Phantom Image Test
Operating Levels
Objective: The objective of this article is to
provide help to facilities that perform screen-film (S-F)
mammography for initially establishing the operating levels for
their phantom image test and for changing the operating levels in a
“for cause” situation.
Note: The following suggestions
apply to facilities using S-F mammography units. For full-field
digital mammography (FFDM) units, facilities should refer to the
unit’s quality control manual.
Establishing operating levels for the first time
Prior to performing mammography with new equipment for the first
time, your medical physicist (MP) must evaluate your mammography
unit and film processor. During or immediately following the
mammography equipment evaluation, when the mammography unit and
processor are operating under optimal conditions, your MP should be
able to assist you in establishing baseline operating levels for the
background density on your phantom images. He/she may recommend an
exposure technique that will result in:
- a background density of 1.80 or higher, and
- a density difference (contrast) with which you can expect to
get a minimum of 0.40 under optimized technique factors and
processing conditions.
Typically, most current mammography S-F combinations are designed
to produce a density difference value of approximately 0.50 or
higher when the background density is approximately 1.80. If the
density difference value on your phantom image is below 0.40, you
should contact your MP regarding how to proceed. Using these
baseline values will help in optimizing the image quality of your
mammogram films. Additionally, it is also helpful to determine
baseline values from phantom images produced over a five-day period;
that will help stabilize the operational levels. An average of the
values from the five phantom images should produce the most
appropriate operating limits.
Once you start performing mammography after setting your baseline
operating levels for background density and density difference, you
should ask your lead interpreting physician whether the
density and contrast of the mammograms are
acceptable. If the interpreting physician feels the mammograms are
of adequate quality, your operating levels are sufficient and you
should continue to use them. However, if your lead interpreting
physician recommends changes, you may want to contact your MP for
help in determining new operating levels.
Changing operating levels for background density and
density difference
Once you have set your operating levels, you should keep them the
same unless your lead interpreting physician suggests changes to
achieve a difference in density or contrast. Any significant changes
occurring in the equipment might also require you to reevaluate the
operating levels. Some examples that would lead you to consider
reestablishing your existing operating levels are:
- Lead interpreting physician decision to increase clinical
image optical density
- The replacement of the mammography x-ray unit or film
processor
- Changes in manufacturer or type of film, screen, or
processing chemistry
Note: You should not reestablish
the operating levels on the recommendation of processor maintenance
personnel without consulting with your MP, as the
MP is responsible for the technical quality of the image.
Before considering a change in operating levels due to
replacement of the mammography x-ray unit or the film processor,
consult with your MP about keeping the same operating levels. A
small change in system settings, such as the density control, may
allow you to continue to use your present operating levels. If your
MP recommends new operating levels, inform your lead interpreting
physician of the change. If new operating levels are established,
ask him/her whether the new levels are producing an acceptable
density and contrast for quality mammograms.
If you change the manufacturer or type of film, screen, and/or
processor chemistry, you may need to set new operating levels. If
you are considering a change in film or screens, you should contact
your MP before making the change. During your
discussion, your MP may recommend a change in operating levels for
your new S-F combination that will work better with a higher
background density and density difference. As mentioned earlier, you
should always notify your lead interpreting physician about any
intended change(s), and immediately following a change, confer with
your interpreting physician regarding the quality of the mammograms.
Cautions about changing operating levels
You should not “automatically” change operating levels when
problems arise with any of your mammography equipment. If the
background density or the density difference start to fluctuate or
drift toward or exceed the operating level action limits, contact
your MP for help in determining the source of this change. If the
change is caused by a problem with a component of your imaging
system, you need to fix the problem so that your facility can
perform high quality mammography. Adjusting or repairing equipment
is a much better approach than frequent changes in the operating
levels to accommodate a problem that could adversely affect the
quality your facility’s mammograms. When you consider changing
operating levels, you should be sure that the change is best for
quality mammograms.
The following Question was submitted to the ACR on 2/21/06:Prompted by a facility question of how & when to
re-establish phantom aims, my question to the ACR: I have
not been able to find guidance on establishing new aims for
phantom controls in neither the ACR manual or the FDA Regs.
I can find reasons why to establish new Phantom aims but not
how. Some facilities are doing multiple day averages and
some are taking the 1st phantom as their new aims. Can you
help?
ACR Response:
Email response from:
Diane A. Zawacki, R.T.(R)(M)(QM)
American College of Radiology
Program Specialist
Breast Imaging Accreditation
Phone: (800) 227-6440 ext. 4948
Fax: (703) 648-9176
Neither the ACR nor FDA have guidance on "re-establishing"
phantom aims. The ACR frequently receives telephone calls
from technologists regarding this very same issue. We think
the technologists are confusing this with re-establishing
processor QC operating levels.
The ACR follows the information provided in the FDA's Policy
Guidance Help System. Please share this information with
your facilities. They will be most interested in #4 and #12
below.
From the Policy Guidance Help System:
Citation:
900.12(e)(2)(i),(ii),(iii),(iv):
Question 4: Under what circumstances should I establish a
new baseline optical density (OD) operating level?
A new baseline OD operating level may need to be established
when switching to a new type of film or if the AEC density
selector settings in the mammographic unit have been
re-calibrated during servicing of the unit. Before changing
operating levels, check that there isn't an underlying
problem that needs to be corrected. A new level may be
established if the interpreting physician(s) has/have made
an intentional decision to modify background optical
densities of the clinical images. For example, many
facilities are choosing to increase film density to take
advantage of the film's increased contrast at higher ODs.
Question 12: If the optical density (OD) for the weekly
phantom test falls below 1.20 (and/or changes by more than
± 0.20 from the established operating level), must the unit
be recalibrated or can we adjust the density setting to
obtain a 1.20 OD?
If the OD at the center of the phantom image falls below the
required minimum of 1.20 (and/or changes by more than ± 0.20
from the established operating level), the facility should
follow pathway A, B, or C; below, based on the situation at
the facility:
-
If the film is of a different type (e.g., switch from
Min-R 2000 to Min-R E) from the previous week's passing
test, the facility should establish new phantom QC operating
levels.
-
If the film emulsion batch is unchanged from the previous
week's passing test:
-
Ensure that the phantom is exposed using typical clinical
conditions and that the position of the phantom and, where
appropriate, the position of the AEC detector have not
changed from that used for prior images.
-
Reevaluate the daily processor performance and make sure
the processor is properly optimized according to the film
manufacturer's specifications.
-
If the facility has been tracking mAs, check the function
of the mammography unit by comparing the mammography unit's
current mAs output with values obtained for previous phantom
images. If the mAs has changed by more than 15%, and the
facility has been using the same kVp, the same mammography
unit density setting, and the processor is operating within
its action limits, then the medical physicist should be
called to check the entire imaging chain, including the
mammography unit. If the mAs has not changed by more than
15% then proceed with step 4. If the facility has not been tracking mAs, the
facility should consult with its medical physicist for what
to do next.
-
If no problems are found in steps 2 and 3, adjust the
density control setting to obtain an optical density of at
least 1.20 at the center of the phantom image (or obtain an
optical density within ± 0.20 of the established operating
level).
-
Adjust the density control setting used clinically to be
consistent with the changes made in step 4.
-
If the film is of the same type but of a different
emulsion batch from the previous week's passing test, the
facility should follow the steps as described in B 1 through
5.
If the optical density again falls below 1.20 (and/or
changes by more than ± 0.20 from the established operating
level) the next time the weekly phantom test is performed,
the facility should follow the appropriate pathway (based on
the film emulsion used) from the following three options:
a. If film of a different type (e.g., switch from Min-R 2000
to Min-R E) is used, the facility should establish new
phantom QC operating levels.
b. If film of the same emulsion batch is used (assuming the
same kVp and mammography unit density settings are used, and
the processor is operating within its action limits), the
facility should consult with its physicist and check the
entire imaging chain before performing mammograms.
c. If film of the same type (but not of the same emulsion
batch) is used, the facility should repeat steps B 1 through
5.
Mammography Home |